GSO IEC 61847:2015
IEC 61847:1998
Gulf Standard
Current Edition
·
Approved on
03 September 2015
Ultrasonics - Surgical systems - Measurement and declaration of the basic output characteristics
GSO IEC 61847:2015 Files
English
31 Pages
Current Edition
Reference Language
Obtaining this standard through the store is currently unavailable. You can acquire it directly from its source.
GSO IEC 61847:2015 Scope
This International Standard specifies:
– the essential non-thermal output characteristics of ultrasonic surgical units;
NOTE 1 – One of the parameters of interest is output acoustic power. This standard addresses only the lowfrequency
(under 100 kHz) component of the total delivered energy. The high-frequency component, which probably
relates to cavitation developed at the tip, is not addressed (see A.4).
– methods of measurement of these output characteristics;
– those characteristics which should be declared by the manufacturers of such equipment.
NOTE 2 – In the interest of clarity, this standard does not address all of the complex surfaces and shapes possible
for applicator tips. A straight tubular shape is used in the description of the parameters and measurements to be
made. It is left to the user of this standard to adapt the basic methodology described to more complex designs if
required.
This International Standard is applicable to equipment which meets the requirements of a, b
and c below:
a) ultrasonic surgical systems operating in the frequency range 20 kHz to 60 kHz; and
b) ultrasonic surgical systems, whose use is the fragmentation or cutting of human tissue,
whether or not those effects are delivered in conjunction with tissue removal or coagulation;
and
c) ultrasonic surgical systems, in which an acoustic wave is conducted by means of a
specifically designed wave guide to deliver energy to the surgical site.
NOTE 3 – Examples of these types of systems are surgical aspirators, intracorporeal lithotripters, end-cutting
devices etc.
This International Standard is not applicable to:
– lithotripsy equipment which uses extracorporeally induced pressure pulses, focussed
through liquid conducting media and the soft tissues of the body;
– surgical devices used as part of the therapeutic process (hyperthermia systems);
– surgical devices whose acoustic application areas are not at the end of a longitudinally
vibrating applicator tip and therefore would not fit the monopole model used in this standard.
This International Standard does not deal with the effectiveness or safety of ultrasonic surgical
systems.
NOTE 4 – Throughout this standard, the term accuracy means the overall uncertainty expressed at the 95%confidence level.
Best Sellers From Electrical Sector
GSO 2530:2016
Gulf Standard
Energy Labelling And Minimum Energy Performance Requirements For Air-Conditioners

OS GSO 2530:2016
GSO 2530:2016
Omani Standard
Energy Labelling And Minimum Energy Performance Requirements For Air-Conditioners


OS GSO 34:2007
GSO 34:2007
Omani Standard
LEAD-ACID STARTER BATTERIES USED FOR
MOTOR VEHICLES AND INTERNAL
COMBUSTION ENGINES


GSO 34:2007
Gulf Technical Regulation
LEAD-ACID STARTER BATTERIES USED FOR
MOTOR VEHICLES AND INTERNAL
COMBUSTION ENGINES

Recently Published from Electrical Sector
GSO 2770:2024
Gulf Standard
WATER HEATERS - ENERGY PERFORMANCE REQUIREMENTS

GSO IEC TS 62257-4:2024
IEC TS 62257-4:2015
Gulf Standard
Recommendations for renewable energy and hybrid systems for rural electrification - Part 4: System selection and design


GSO IEC 61968-13:2024
IEC 61968-13:2021
Gulf Standard
Application integration at electric utilities - System interfaces for distribution management - Part 13: Common distribution power system model profiles


GSO IEC 62246-4:2024
IEC 62246-4:2023
Gulf Standard
Reed switches – Part 4: Application in conjunction with magnetic actuator used for magnetic sensing devices