GSO IEC 60601-2-68:2017

IEC 60601-2-68:2014
Gulf Standard   Current Edition · Approved on 03 October 2017

Electrical medical equipment - Part 2-68: Particular requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam

GSO IEC 60601-2-68:2017 Files

GSO IEC 60601-2-68:2017 Scope

Replacement: This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of X-ray based IMAGE-GUIDED RADIOTHERAPY equipment for use with EXTERNAL BEAM EQUIPMENT (EBE). This particular standard covers safety aspects of kilovoltage (kV) and megavoltage (MV) X-ray imaging devices in a known geometrical relationship with EBE for the purpose of IGRT. It covers aspects of communication and relationships between the EXTERNAL BEAM EQUIPMENT and X-ray imaging devices, attached or not directly attached to, but in the same RADIATION shielded area as, and dedicated for use only with the EXTERNAL BEAM EQUIPMENT. This particular standard deals with equipment for REAL-TIME X-IGRT, ONLINE X-IGRT and OFFLINE X-IGRT. It covers procedures to reduce the risk of over-reliance on the X-IGRT EXTERNAL BEAM SYSTEM (X-IGRT EBS). For example the manufacturer will provide an interactive interface for user interaction with the correction suggested by the system. If a clause or subclause is specifically intended to be applicable to X-IGRT EBE SYSTEMS the content of that clause or subclause will say so. If that is not the case, the clause or subclause applies only to X-IGRT EQUIPMENT. This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the MANUFACTURER and some installation aspects of X-IGRT EBE SYSTEMS intended to be • for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular specified clinical purposes, e.g. STATIONARY RADIOTHERAPY or MOVING BEAM RADIOTHERAPY • maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE, • subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON. NOTE In this particular standard, all references to installation refer to installation in the RESPONSIBLE ORGANIZATION’S premises

Best Sellers From Electrical Sector

GSO 2530:2016
 
Gulf Standard
Energy Labelling And Minimum Energy Performance Requirements For Air-Conditioners
OS GSO 2530:2016
GSO 2530:2016 
Omani Standard
Energy Labelling And Minimum Energy Performance Requirements For Air-Conditioners
OS GSO 34:2007
GSO 34:2007 
Omani Standard
LEAD-ACID STARTER BATTERIES USED FOR MOTOR VEHICLES AND INTERNAL COMBUSTION ENGINES
GSO 34:2007
 
Gulf Technical Regulation
LEAD-ACID STARTER BATTERIES USED FOR MOTOR VEHICLES AND INTERNAL COMBUSTION ENGINES

Recently Published from Electrical Sector

GSO IEC 61558-2-2:2024
IEC 61558-2-2:2022 
Gulf Standard
Safety of transformers, reactors, power supply units and combinations thereof - Part 2-2: Particular requirements and tests for control transformers and power supply units incorporating control transformers
GSO IEC 60664-1:2024
IEC 60664-1:2020 
Gulf Standard
Insulation coordination for equipment within low-voltage supply systems - Part 1: Principles, requirements and tests
GSO IEC 61936-2:2024
IEC 61936-2:2023 
Gulf Standard
Power installations exceeding 1 kV AC and 1,5 kV DC - Part 2: DC
GSO IEC TS 62915:2024
IEC TS 62915:2023 
Gulf Standard
Photovoltaic (PV) modules - Type approval, design and safety qualification - Retesting