GSO ISO 12846:2013
ISO 12846:2012
Gulf Standard
Current Edition
·
Approved on
31 December 2013
Water quality -- Determination of mercury -- Method using atomic absorption spectrometry (AAS) with and without enrichment
GSO ISO 12846:2013 Files
English
15 Pages
Current Edition
Reference Language
OMR
46.24
GSO ISO 12846:2013 Scope
This International Standard specifies two methods for the determination of mercury in drinking, surface, ground, rain and waste water after appropriate pre-digestion. For the first method (described in Clause 6), an enrichment step by amalgamation of the Hg on, for example, a gold/platinum adsorber is used. For the method given in Clause 7, the enrichment step is omitted.
The choice of method depends on the equipment available, the matrix and the concentration range of interest. Both methods are suitable for the determination of mercury in water. The method with enrichment (see Clause 6) commonly has a practical working range from 0,01 µg/l to 1 µg/l. The mean limit of quantification (LOQ) reported by the participants of the validation trial (see Annex A) was 0,008 µg/l. This information on the LOQ gives the user of this International Standard an orientation and does not replace the estimation of performance data based on laboratory-specific data. It has to be considered that it is possible to achieve lower LOQs with specific instrumentation (e.g. single mercury analysers).
The method without enrichment (in Clause 7) commonly has a practical working range starting at 0,05 µg/l. The LOQ reported by the participants of the validation trial (see Annex A) was 0,024 µg/l. It is up to the user, based on the specific application, to decide whether higher concentrations are determined by omitting the enrichment step and/or by diluting the sample(s). The sensitivity of both methods is dependent on the selected operating conditions.
Another possibility for the determination of extremely low Hg concentrations down to 0,002 µg/l without pre-concentration is the application of atomic fluorescence spectrometry (see ISO 17852). Specific atomicabsorption mercury analysers allow determinations down to 0,010 µg/l without pre-concentration.
In general, the determination of trace concentrations of Hg by AAS (or AFS) is dependent on clean operating conditions in the laboratory and on the use of high-purity chemicals with negligible low-Hg blanks.
NOTE This International Standard may be applied to industrial and municipal waste water after an additional digestion step performed under appropriate conditions and after suitable method validation (see 7.4). A potential sample stability issue (mercury loss) for anaerobic reducing industrial effluents has to be considered thoroughly.
Best Sellers From Chemical and Textile Sector
OS GSO 1943:2024
GSO 1943:2024
Omani Standard
Cosmetic Products – Safety Requirements of Cosmetics and Personal Care Products
GSO 1943:2024
Gulf Technical Regulation
Cosmetic Products – Safety Requirements of Cosmetics and Personal Care Products
OS GSO 2528:2024
GSO 2528:2024
Omani Standard
Cosmetic products – Technical Regulation of Cosmetic and Personal Care Products Claims
GSO 2528:2024
Gulf Technical Regulation
Cosmetic products – Technical Regulation of Cosmetic and Personal Care Products Claims
Recently Published from Chemical and Textile Sector
OS GSO ASTM E2752:2026
ASTM E2752:25
Omani Standard
Standard Guide for Evaluation of Residual Effectiveness of Antibacterial Personal Cleansing Products
OS GSO 2815:2026
GSO 2815:2026
Omani Standard
Cosmetic products – Guidelines on labeling requirements of fragrance allergens
OS GSO EN 16274:2026
EN 16274:2021
Omani Standard
Method for analysis of allergens - Quantification of an extended list of 57 suspected allergens in ready to inject fragrance materials by gas chromatography mass spectrometry
GSO ASTM E2752:2026
ASTM E2752:25
Gulf Standard
Standard Guide for Evaluation of Residual Effectiveness of Antibacterial Personal Cleansing Products