OS GSO ISO 4965-1:2021

ISO 4965-1:2012
Omani Standard   Current Edition · Approved on 01 July 2021

Metallic materials — Dynamic force calibration for uniaxial fatigue testing — Part 1: Testing systems

OS GSO ISO 4965-1:2021 Files

OS GSO ISO 4965-1:2021 Scope

ISO 4965-1:2012 describes two methods for determining the relationship between the dynamic force range applied to a test-piece in a uniaxial, sinusoidal, constant amplitude test and the force range indicated by the testing system.

These methods are applicable to dynamic testing systems operating away from system resonant frequencies and are relevant to testing systems where the dynamic force measurement errors are either unknown or where they are expected to exceed 1 % of the applied force range.

The dynamic force measurement errors are determined by comparison of the peak forces indicated by the dynamic testing system with those measured by the strain gauged dynamic calibration device (DCD). This DCD has previously undergone static calibration against the testing system indicator.

For Method A (Replica test-piece method), the dynamic calibration is applicable over the validated range of frequencies for that type of test-piece only. A frequency-dependent correction factor is applicable for the correction of dynamic force measurement errors of up to 10 % of dynamic force range. By using such a correction factor, the actual test specimen dynamic force measurement error will be reduced to less than 1 % of the dynamic force range.

For Method B (Compliance envelope method), the dynamic calibration is applicable over the range of test frequencies validated for test-pieces whose compliance lies between those of the two DCDs. No correction factor is applicable, as Method B does not permit dynamic force measurement errors above 1 % of the dynamic force range.

Best Sellers From Metrology Sector

OS GSO OIML R87:2021
OIML R87:2016 
Omani Standard
Quantity of product in prepackages
GSO OIML R87:2021
OIML R87:2016 
Gulf Standard
Quantity of product in prepackages
GSO OIML R79:2021
OIML R79:2015 
Gulf Standard
Labeling requirements for prepackages
OS GSO OIML R79:2021
OIML R79:2015 
Omani Standard
Labeling requirements for prepackages

Recently Published from Metrology Sector

GSO ISO 10713:2026
ISO 10713:2025 
Gulf Standard
Jewellery and precious metals — Gold alloy coatings
GSO ISO 20816-2:2026
ISO 20816-2:2017 
Gulf Standard
Mechanical vibration — Measurement and evaluation of machine vibration — Part 2: Land-based gas turbines, steam turbines and generators in excess of 40 MW, with fluid-film bearings and rated speeds of 1 500 r/min, 1 800 r/min, 3 000 r/min and 3 600 r/min
GSO ISO 23783-2:2026
ISO 23783-2:2022 
Gulf Standard
Automated liquid handling systems — Part 2: Measurement procedures for the determination of volumetric performance
GSO ISO 16610-45:2026
ISO 16610-45:2025 
Gulf Standard
Geometrical product specifications (GPS) — Filtration — Part 45: Morphological profile filters: Segmentation